WRIGHT MEDICAL TECHNOLOGY INC INFINITY ADAPTIS TIB SZ3LNG INFINITY ADAPTIS; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Catalog Number |
33680013 |
| Medical Device Problem Codes |
Osseointegration Problem (3003); Implant subsidence (4078)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
| Date of Event |
08/20/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The stryker prophecy team has received a ct scan for upcoming revision surgery.The reason for the revision is due to subsidence of the tibial component, malalignment, and migration.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information has been requested and if received, will be provided in the supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction - please refer to d4 lot#.The reported event could be confirmed since images of ct scans were provided and shows loosening and migration of the tibial and talar components.Upon further investigation of the ct scans by healthcare professionals the following was observed: there is a clear and massive loosening and migration of both the tibial and the talar component.The tibia is almost luxated from the joint space and should be revised as soon as possible.The underlying cause is not clear, but it could definitely be related to a poor bone substance and therefore patient related factor.Failure of total ankle replacement tar over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient radiological and clinical information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, this assessment is limited.No conclusive statement can be provided, because key clinical information e.G., clinical status, exact symptoms, and range of motion of the patient, assessment of the treating physician is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Due to these limitations, we are unable in such cases to provide statements about the patient, the procedure and or the device in relation to cause of the failure.Although the issue is most likely related to poor bone substance, the root cause could not be conclusively determined.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
The stryker prophecy team has received a ct scan for upcoming revision surgery.The reason for the revision is due to subsidence of the tibial component, malalignment, and migration.
|
| |
|