• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SRM-560
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 07/09/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade ii.
 
Event or Problem Description
Healthcare professional reported left side "no complaint against the device via mammogram and mri".Healthcare professional later reported "rupture, and suspected capsular contracture baker grade ii".Healthcare professional additionally reported "left side was not confirm to be rupture, found to be intact during the surgery".Device has been explanted, replaced, and discarded.
 
Additional Manufacturer Narrative
Device analysis results: based on the product analysis performed, the assessments of the complaints are: ¿ capsular contracture: unable to observe since it is a medical event and is not related to the device.As per the investigation procedure, deformation and crease were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
 
Event or Problem Description
Healthcare professional reported left side "no complaint against the device via mammogram and mri".Healthcare professional later reported "rupture, and suspected capsular contracture baker grade ii".Healthcare professional additionally reported "left side was not confirm to be rupture, found to be intact during the surgery".Device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key23058797
Report Number9617229-2025-15640
Device Sequence Number21003929
Product Code FTR
UDI-Device Identifier10888628007178
UDI-Public(01)10888628007178(10)3091194(11)170814(17)220714
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/14/2022
Device Catalogue NumberSRM-560
Device Lot Number3091194
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/28/2025
Supplement Date Received by Manufacturer10/20/2025
Initial Report FDA Received Date09/15/2025
Supplement Report FDA Received Date11/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/14/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
Patient SexFemale
Patient Weight73 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-