• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS MIDAS REX MR8; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS MIDAS REX MR8; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE Back to Search Results
Model Number MR8-LP05BA40D
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis: no conclusion can be drawn.No evaluation was performed, as the device was discarded by the customer.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the tool has broken.No patient impact reported.On follow up, it was reported that there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX MR8
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key23060039
Report Number1625507-2025-02064
Device Sequence Number21250376
Product Code HBE
UDI-Device Identifier00763000513580
UDI-Public(01)00763000513580(17)280405(10)0228631289(20)01
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K183515
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR8-LP05BA40D
Device Catalogue NumberMR8-LP05BA40D
Device Lot Number0228631289
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2025
Initial Report FDA Received Date09/16/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-