| Catalog Number |
63612 |
| Medical Device Problem Code |
Sharp Edges (4013)
|
| Health Effect - Clinical Code |
Abrasion (1689)
|
| Date of Event |
09/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient advised that they noticed the packaging changed on the last order and the patient said they checked the codes and the same, so they used them but found some differences.The patient said the individual sleeves were crammed together and hard to separate.Patient also said the edges of the sleeves were very sharp and they got a few cuts from them.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.Instructions for use reviewed for this investigation.The reported event is addressed within the labeling as it state" precautions: do not use device if package is opened or damaged.Inspect the catheter before use.Do not use the device if damaged.If you have any clinical concerns with the use of this device for your condition, please contact your healthcare physician.Users and/or patients within the european union, should report any serious incident that has occurred in relation to the device to the manufacturer and the competent authority of the member state in which the user and/or patient is established.Users outside of the european union should report any serious incident that has occurred in relation to the device to the manufacturer and the regulatory authority of the country in which the user and/or patient is established." a review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Corrections: d, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient advised that they noticed the packaging changed on the last order and the patient said they checked the codes and the same, so they used them but found some differences.The patient said the individual sleeves were crammed together and hard to separate.Patient also said the edges of the sleeves were very sharp and they got a few cuts from them.
|
| |
|
Search Alerts/Recalls
|