|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause of the reported failure can be attributed to user error of the device due to prying/leveraging the attachment during use.Per (b)(4).Do not allow the rotating portion of any device to touch any metallic object, such as a cannula or arthroscope, during use.Damage to both devices is likely.Damage to the device can range from a slight distortion or dulling of the blade/burr edge to actual fracture of the tip in vivo.If such contact does occur, inspect the tip.If there are cracks, fractures or dulling, or if there is any other reason to suspect a blade is damaged, replace it immediately.Excessive "side-loading" on the device during use does not improve cutting performance and in extreme cases will cause irreversible damage to the device.
|