DEPUY MITEK LLC US GRYPHON T BR ANCHOR W/OC; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
210812 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: the product has not returned to j&j medtech orthopaedics, however a photo was provided for evaluation.¿ visual inspection of the returned photo found the anchor still attached to the inserter and broken at the distal end, the anchor had foreign matter at the distal end.No other anomalies could be observed.The overall complaint was confirmed as the observed condition of the gryphon t br anchor w/oc would have contributed to the complained device issue.Device history review: a review of the device history record was performed which indicated that there were no relevant issues identified during the manufacture of the device that may have contributed to the reported condition.¿ based on the investigation findings, the potential cause for the reported condition is traced to the procedural variables, such handling of the device or product interaction during procedure.The possible root cause can be associated with off axis insertion and levering during insertion.As per the instructions for use, axial misalignment or levering with the anchor upon insertion, may result in anchor or inserter fracture.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Event or Problem Description
|
|
It was reported that during a ankle joint ligament repair surgery, the gryphon t br anchor w/oc was broken off.It was reported that all broken parts were removed.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.H.11.Additional narrative: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection found that the device had the anchor still attached to the shaft and broken at the distal end.Not all the broken fragments were returned for evaluation.One suture was detached from the inserter and had a fragment of the anchor attached to it.The device had foreign matter, presumably biological matter.No other anomalies were noted.The overall complaint was confirmed as the observed condition of gryphon t br anchor w/oc would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to the procedural variables, such handling of the device or product interaction during procedure.The possible root cause can be associated with off axis insertion and levering during insertion and sub-consequently led the anchor to break/damage.As per the instructions for use, 1) axial misalignment or levering with the anchor upon insertion, may result in anchor or inserter fracture.2) apply tension (normal force) on suture lengths, as excessive force may overload the anchor or suture.Properly handling and attention to the approved use of the device diminishes the risk of failure.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: a review of the device history record was performed which indicated that there were no relevant issues identified during the manufacture of the device that may have contributed to the reported condition.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: d4.Expiration date; d4.Primary udi number; h4.Device manufacture date: updated accordingly.
|
| |
|