|
Updated field(s): d8, h3, h4, h6, h1.This supplemental report is being submitted to provide the results of the final investigation.The device was returned to olympus for inspection, and the reported issue was confirmed.Based on the results of the investigation, a definitive root cause could not be identified, however, it is possible that a high-energy treatment device (such as high-frequency device) was used without ensuring a sufficient distance from the tip.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|