| Model Number |
9445-24 |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a signal loss occurred.It was indicated the patient started their transmitter past the 'use by' date, which is misuse of the device.No product or data was provided for evaluation.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that signal loss occurred.It was indicated the patient started their transmitter past the 'use by' date, which is misuse of the device.Product was provided for investigation.An external visual inspection was performed and passed.Voltage test was performed and passed.Nordic bluetooth pairing test was performed and failed.Performance data was reviewed.The allegation was confirmed due to the finding of signal loss over an hour within the investigation window.The probable cause was a defective transmitter.No injury or medical intervention was reported.
|
| |
|