Per the instructions for use (ifu), structural valve deterioration (wear, fracture, calcification, leaflet tear/tearing from the stent posts, leaflet retraction, suture line disruption of components of a prosthetic valve, thickening, stenosis), and device degeneration are known potential risks associated with the use of bioprosthetic heart valves.The ifu cautions that accelerated deterioration of the valve may occur in patients with an altered calcium metabolism.Long-term durability has not been established for the valve.Regular medical follow-up is advised to evaluate valve performance.Structural valve deterioration (svd) may be manifested as stenosis with thickened leaflets.Svd may be mild and not require any intervention, or it may be moderate to severe.It can cause the heart to work harder to eject blood from the ventricle.Depending on the severity it could be an indication for valve replacement or medical intervention.Tissue calcification is a common failure mode.The mechanisms for bioprosthetic heart valve tissue calcification are not yet fully understood.Many factors can contribute to the onset and propagation of calcification including patient-related (e.G., patient age, disease state, immune status, and other co-morbidities), pharmacological, and intrinsic properties of the valve itself.It is widely understood that patients with chronic renal disease and prior history of calcific stenosis of the native valve may be predisposed to bioprosthetic calcification.During the manufacturing process, all sapien thv valves are 100% visually inspected for defects and 100% functionally tested for proper coaptation under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.In this case, there was no allegation or indication a manufacturing non-conformance contributed to this product problem.Due to the limited information provided, the cause of the degeneration is unknown.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
|
|
As reported by the field clinical specialist, approximately 7 years post transfemoral transcatheter pulmonic valve replacement (tpvr) procedure with a 23 mm sapien xt valve, the valve was failing due to regurgitation and moderate to severe stenosis.A catheterization revealed a pre-procedural pressure gradient of >35 mmhg.The initial balloon dilations were performed using a 24 mm x 4 cm non-edwards balloon, followed by a 26 mm x 4 cm non-edwards balloon, which showed a small waist.The sapien xt valve was described as ''napkin ringed.'' a valve-in-valve procedure was performed using a 26 mm sapien 3 valve, implanted with an overfill of +3cc.A visual gap between the valve frames was noted, prompting a post-dilation with an additional +1cc, which resulted in minimal improvement.The pressure gradient decreased to approximately 15 mmhg.A final balloon inflation using a 26 mm x 4 cm non-edwards balloon at 12 atm further reduced the gradient to around 10 mmhg.Intracardiac echocardiography (ice) confirmed that there was no residual leak.No patient complications were reported.
|
|
A supplemental mdr is being submitted due to additional information from medical record review.The following sections of this report have been updated: b.4, b.5, b.7 g.3, g.6, h.2 and h.6 clinical code corrected.
|