MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK/CONTINU-FLO; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK/CONTINU-FLO; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 2C8571
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a clearlink system continu-flo solution set leaked from the y-site.This was observed prior to patient use.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A pressure test and clear passage under water were performed, and a leak in the first y-site was observed.An autopsy was performed in the second y-site clearlink, and a hold in the gland was observed.The reported condition was verified.The cause was determined to be related to manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK/CONTINU-FLO
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
parque industrial de cartago
apartado no. 1-7052
cartago
CS  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key23069092
Report Number1416980-2025-04634
Device Sequence Number9808988
Product Code FPA
UDI-Device Identifier00085412046327
UDI-Public(01)00085412046327(10)R25C26062
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K153158
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2C8571
Device Catalogue Number2C8571
Device Lot NumberR25C26062
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2025
Initial Report FDA Received Date09/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-