• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EVOQUE TRICUSPID VALVE REPLACEMENT SYSTEM; PERCUTANEOUSLY DELIVERED PROSTHESES AND TRICUSPID VALVES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EVOQUE TRICUSPID VALVE REPLACEMENT SYSTEM; PERCUTANEOUSLY DELIVERED PROSTHESES AND TRICUSPID VALVES Back to Search Results
Model Number 9850EV48US
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Heart Block (4444)
Date of Event 09/02/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The manufacturer's investigation is ongoing and a follow-up report will be submitted when the investigation is complete.
 
Event or Problem Description
Edwards received notification of a transcatheter tricuspid valve replacement (ttvr) procedure involving the evoque system where upon deployment, patient went into complete heart block.The patient was paced off wire and a temp pacer was placed through the valve.Ep was consulted and they placed a temp pacer lead from the ij.The patient is currently pacer dependent and in stable condition.The conduction disturbance was noticed right at deployment.The final valve deployment position was good.Per medical opinion, the perceived root cause of the event was the valve interacting with the conduction system.
 
Additional Manufacturer Narrative
The following sections were updated/corrected/added: b4, g3, g6, h2, h6, and h11.The complaint for valve interacts with conduction system was confirmed with other empirical evidence via information received by edwards.Based on the thorough investigation completed during the complaint reviews, no product issues have been identified.The investigations did not reveal any evidence of a manufacturing defect or device-related issue that could have contributed to the events.Per the instructions for use (ifu), conduction system disturbance and/or heart block which may require treatment (including permanent pacemaker) are potential adverse events.Heart block and other conduction disturbances are common in patients with underlying cardiovascular disease and can be exacerbated or aggravated with standard intra-operative medications or anesthesia.Such events are related to the patient's underlying condition.Other mechanisms for conduction abnormalities could be related to coronary obstruction due to air embolism, coronary spasm and/or local edema affecting the av node.Conduction disturbances have also been documented during transcatheter cardiac procedures as a result of direct interaction with cardiac anatomy, especially in patients with underlying conduction abnormalities.Other potential causes include trauma by a guidewire or delivery system.Additionally, cardiac conduction system complications are known risks with tricuspid valve interventions due to the proximity of the atrioventricular node (av node) to the septal leaflet of the tricuspid valve.These conduction disturbances can lead to post-operative heart block requiring pacemaker implantation.Nevertheless, a definitive root cause cannot be determined.Since there are no confirmed product or labeling/ifu/training material non-conformances affecting distributed product and no other triggers have been met, no corrective or preventative actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOQUE TRICUSPID VALVE REPLACEMENT SYSTEM
Common Device Name
PERCUTANEOUSLY DELIVERED PROSTHESES AND TRICUSPID VALVES
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
cassandra cook
1 edwards way
irvine, CA 92614
5743778277
MDR Report Key23069127
Report Number2015691-2025-07638
Device Sequence Number10044949
Product Code NPW
UDI-Device Identifier00690103217452
UDI-Public(01)00690103217452(17)270423(11)250423
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P230013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number9850EV48US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2025
Supplement Date Received by Manufacturer10/23/2025
Initial Report FDA Received Date09/16/2025
Supplement Report FDA Received Date10/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-