• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE I/A HANDPIECE TIP; CATHETER, IRRIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE I/A HANDPIECE TIP; CATHETER, IRRIGATION Back to Search Results
Catalog Number 8065752066
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
A physician reported that the ultrasound tip (phaco emulsification tip) was too bent to use before surgery.The surgery was performed and completed after replacing the product with another one.The procedure type was unknown.There was no information about any patient impact.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.No technical inquiries were received from the customer.One opened phaco tip in a tip wrench was received for the report.The returned sample was visually inspected and was found to be nonconforming with phaco emulsification tip (phaco tip) cannula bent above the aspiration bypass hole, no wear was observed on threads, back of flange, and nut corners and consistent with threading on handpiece.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a nonconformance review of the reported lot.A nonconformance was found.This non-conformance could have potentially contributed to the reported event.An investigation was initiated in order to investigate the root cause of the bent phaco tip.The returned sample was found to be visually nonconforming with a bent phaco tip.How and when the phaco tip became bent could not be determined due to that the tip was loose in the bag without the protective wrench.However, because no wear was found on the threads a potential manufacturing related root cause could not be ruled out.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.As the root cause and its origin are inconclusive, further investigation or manufacturing actions are not warranted at this time.However, an investigation has been initiated to determine root cause of the bent phaco tip.Additionally, nonconformance investigation was initiated for the related nonconformance identified on the nonconformance report.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Any nonconformance is removed from the lot and scrapped.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I/A HANDPIECE TIP
Common Device Name
CATHETER, IRRIGATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23070677
Report Number2523835-2025-00987
Device Sequence Number18641236
Product Code GBX
UDI-Device Identifier00380657520664
UDI-Public(01)00380657520664(17)291231(11)250124(10)177W58
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K910245
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065752066
Device Lot Number177W58
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/26/2025
Supplement Date Received by Manufacturer02/09/2026
Initial Report FDA Received Date09/17/2025
Supplement Report FDA Received Date03/05/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/24/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-