Additional information provided in d.9., h.3., h.6.And h.11.No technical inquiries were received from the customer.One opened phaco tip in a tip wrench was received for the report.The returned sample was visually inspected and was found to be nonconforming with phaco emulsification tip (phaco tip) cannula bent above the aspiration bypass hole, no wear was observed on threads, back of flange, and nut corners and consistent with threading on handpiece.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a nonconformance review of the reported lot.A nonconformance was found.This non-conformance could have potentially contributed to the reported event.An investigation was initiated in order to investigate the root cause of the bent phaco tip.The returned sample was found to be visually nonconforming with a bent phaco tip.How and when the phaco tip became bent could not be determined due to that the tip was loose in the bag without the protective wrench.However, because no wear was found on the threads a potential manufacturing related root cause could not be ruled out.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.As the root cause and its origin are inconclusive, further investigation or manufacturing actions are not warranted at this time.However, an investigation has been initiated to determine root cause of the bent phaco tip.Additionally, nonconformance investigation was initiated for the related nonconformance identified on the nonconformance report.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Any nonconformance is removed from the lot and scrapped.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|