| Catalog Number |
8065750853 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A nurse reported that phacoemulsification (phaco) tip broken and fell down in patients¿ eye during the cataract [with intraocular lens (iol) implant] surgery.The surgery details and patient details were unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.6.And h.11.No technical inquiries were received from the customer.A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The attached photos were reviewed by the manufacturing site.The photo 1, shows the pak label, the reported product and lot information is confirmed.The photo 2, show 3 phacoemulsification (phaco) tips with broken tip.Reported issue confirmed from photos.Conformance achieved: the investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.The evaluation of the attached photo confirms the broken phaco tip noted by the customer.Based on the evaluation of the information, since the sample was not received at the manufacturing site, how and when the phaco tip is broken was unable to identify the root cause or origin of the reported event.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information requested and received that the surgery was completed on same day by replacing product.The product was contacted with patient but no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in a.1., a.2., a.3.A., b.5.And h.6.The manufacturer internal reference number is: (b)(4).
|
| |
|