|
Additional information provided in d.9., h.3., h.6.And h.11.No technical inquiries were received from the customer.One probe sample was received for the report; however, the reported lot number was expired when the sample was received at the manufacturing site.Therefore, no product evaluation was performed due to the reported functional issue.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.A sample was returned and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria.However, when the sample was received at the manufacturing site it was determined that the device was beyond its expiry date; therefore, a root cause was unable to be determined.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.A nonconformance investigation has been completed to investigate the trend identified for probes with would not cut or actuate issues.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|