Model Number ND2110X15B
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2024
Type of Reportable Event Malfunction
Event or Problem Description
The trilogy evo ventilator was returned to the third party service center for completion of preventative maintenance.During the evaluation it was noted that the detachable battery and the internal battery were missing from the device.A ventilator service required alarm error code e-46 appeared several times in the device error log indicating the internal battery could be detected.Based on error code e-46, the internal battery was replaced.Additional findings not related to the malfunction/issue: the detachable battery harness was defective requiring replacement.The air foam inlet filter will be replaced due to preventive maintenance.The in-line proximal filter was contaminated and required replacement.The vanity cover had cosmetic damage and required replacement.This device has been serviced, inspected, tested, met acceptance criteria in accordance with all of the applicable customer requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23071756
Report Number2518422-2025-110289
Device Sequence Number4645131
Product Code CBK
UDI-Device Identifier00606959055513
UDI-Public(01)00606959055513
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberND2110X15B
Device Catalogue NumberND2110X15B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2025
Initial Report FDA Received Date09/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/26/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-