Model Number CC60WF
Medical Device Problem Code Failure to Deliver (2338)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
A non-healthcare professional reported that, during an intraocular lens (iol) implantation surgery, the lens was defective, stuck/jammed in cartridge.The procedure was completed using unspecified backup lens.Additional information was requested.
 
Additional Manufacturer Narrative
A product was not returned for analysis.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23072512
Report Number1119421-2025-02489
Device Sequence Number21331243
Product Code HQL
UDI-Device Identifier00380652396868
UDI-Public(01)00380652396868(17)290512(11)250514(10)16036557
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCC60WF
Device Lot Number16036557
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/26/2025
Initial Report FDA Received Date09/17/2025
Date Device Manufactured05/14/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-