| Model Number |
WB991036 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparaskopi procedure, the subject generator device gave an error 274.There was no report of patient harm or injury.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, h2, h3, h6 (type of investigation, investigation findings, investigation conclusions), h11 corrected field: b5 - describe event or problem updated to add the common device name.D2a - common device name updated to electrosurgical system generator.The device was returned to olympus for inspection, and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the error code malfunction: electrical/electronic component problem; expected or random component failure without any design or manufacturing issue.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic procedure, the generator gave an error 274.The procedure was completed with an olympus back-up device.There was no report of patient harm or injury.
|
| |
|