• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ID NOW COVID-19 2.0 TEST KIT 24T (EUA); SIMPLE POINT-OF-CARE DEVICE TO DETECT SAR-COV-2 NUCLEIC ACID TARGETS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ID NOW COVID-19 2.0 TEST KIT 24T (EUA); SIMPLE POINT-OF-CARE DEVICE TO DETECT SAR-COV-2 NUCLEIC ACID TARGETS Back to Search Results
Catalog Number 192-000
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3 (date of event): the date listed is an estimated date, as the consumer did not provide the actual date of the event.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer reported three (3) false negative results with the id now covid-19 assay performed on an unknown date using direct tested throat and nasopharyngeal samples.This is report one (1) of three (3).The patient initially tested negative with the id now covid-19 assay.Repeat testing was performed on two (2) different id now instruments, which generated negative results.Confirmation pcr testing was performed on (b)(6) 2025 with a nasal swab sample that generated a positive result.No additional patient information, including treatment and outcome, was provided.
 
Additional Manufacturer Narrative
B3 (date of event): the date listed is an estimated date, as the consumer did not provide the actual date of the event.Corrections: a3b - gender.D1 - brand name.D2a - common device name, full name: simple point-of-care device to detect sar-cov-2 nucleic acid targets from clinical specimens in near-patient settings.D2b - procode corrected to align with d1.D4 - lot number added, catalog number corrected, expiration date added, serial number made na.D8 - made na.G4 - 510k number added.H4 - manufacture date added h6 - health impact code updated.Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 000m983346 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 192-000 / lot: 000m983346, test base part number 192-430 / lot: 000m983346.The lot met the required release specifications.A review of the complaints reported as false negative patient results (confirmed and unconfirmed, conflicting results) related to kit lot 000m983346 showed that the complaint rate is (b)(4).Abbott diagnostics scarborough, inc.Was unable to determine the exact root cause of the reported issue.
 
Event or Problem Description
The customer reported three (3) false negative results with the id now covid-19 assay performed on an unknown date using direct tested throat and nasopharyngeal samples.This is report one (1) of three (3).The patient initially tested negative with the id now covid-19 assay.Repeat testing was performed on two (2) different id now instruments, which generated negative results.Confirmation pcr testing was performed on (b)(6) 2025 with a nasal swab sample that generated a positive result.No additional patient information, including treatment and outcome, was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ID NOW COVID-19 2.0 TEST KIT 24T (EUA)
Common Device Name
SIMPLE POINT-OF-CARE DEVICE TO DETECT SAR-COV-2 NUCLEIC ACID TARGETS
Manufacturer (Section D)
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
10 southgate road
scarborough ME 04074
Manufacturer Contact
rachel blackwell
10 southgate road
scarborough, ME 04074
6613888803
MDR Report Key23073822
Report Number1221359-2025-00368
Device Sequence Number11958589
Product Code QWR
UDI-Device Identifier00811877011354
UDI-Public(01)00811877011354(17)270122(10)000M983346
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K221925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number192-000
Device Lot Number000M983346
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2025
Supplement Date Received by Manufacturer10/10/2025
Initial Report FDA Received Date09/17/2025
Supplement Report FDA Received Date11/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age50 YR
Patient SexMale
-
-