| Model Number |
19AGN-751 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a 19mm sjm regent heart valve w/ flex cuff was chosen for a procedure.During device preparation, the leaflet function was tested by a leaflet tester and confirmed leaflets where moving normally.During procedure, initially sized the annulus and implanted the valve.After implanting the valve, they found that the leaflet doesn't open properly.So, the surgeon explanted valve and implanted a new 17mm sjm masters series flex cuff hemodynamic plus valve while patient on bypass.Post explant, it was noted that the leaflet was moving.There was no delay in the procedure.The patient was reported stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of incomplete leaflet coaptation was reported.The valve was sent for functional testing at the engineering test lab.During this test, which ensures proper leaflet coaptation and hemodynamic performance, the valve was placed in a pulse duplicator filled with saline.The conditions included a beat rate of 70 ± 1 bpm, a flow rate of 5.0 ± 0.2 lpm, and a mean aortic pressure of 100 +10/-0 mmhg.Under these conditions, the leaflets appeared to coapt such that the effective orifice area value was 1.54 and the regurgitant fraction value was 7.0%.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, including functional testing to ensure proper leaflet coaptation and hemodynamic performance, and the product met all specifications.The cause of reported event could not be determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Search Alerts/Recalls
|