• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB BASE UNIT SERVO-N; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB BASE UNIT SERVO-N; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 6688600
Medical Device Problem Code Tidal Volume Fluctuations (1634)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the ventilator generated alarms for low minute volume.There was no patient harm.Manufacturer's ref #: (b)(4).
 
Additional Manufacturer Narrative
On-site investigation was performed by our field service engineer.The claimed issue was not reproduced during on-site visit.No part was replaced.The ventilator passed all functional and safety tests and was cleared for clinical use.Review of the provided device logs confirmed occurrence of the reported issue.The event log confirms occurrence of several clinical alarms, as an indication of difficulties with ventilating the patient, but there are no technical error codes to indicate a ventilate malfunction.Alarms will be generated when set alarm limits are exceeded, as per design to alert the operator about ongoing issues.Generated alarms are indicating that there is an issue with delivering of the set volumes.The cause of the claimed issue in this customer product complaint has not been determined, as the source of the issue is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE UNIT SERVO-N
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key23073895
Report Number8010042-2025-0001428
Device Sequence Number11923550
Product Code CBK
UDI-Device Identifier07325710001042
UDI-Public(01)07325710001042(11)250225
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K201874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6688600
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2025
Initial Report FDA Received Date09/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/25/2025
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-