Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, the patient underwent an initial surgery with the products in question.After the surgery on (b)(6) 2025, the breakage of the stem calcar was confirmed.The revision surgery was scheduled on (b)(6) 2025.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the device lot number is unknown; therefore, a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.Added: d10 concomitant.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint recon described in this report.H11 additional narrative: added: d9.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary- it was reported that on (b)(6) 2025, the patient underwent an initial surgery with the products in question.After the surgery on (b)(6) 2025, the breakage of the stem calcar was confirmed.The revision surgery was scheduled on sep (b)(6) 2025.No further information is available.The product was returned to depuy synthes for evaluation.Visual analysis of the returned s-rom*sleeve prx ztt, 14b-sml revealed bone ingrowth on the fixation surface.Additionally, marks and scratches were present on the upper section, preventing retrieval of the lot number.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was not confirmed as the observed condition of the s-rom*sleeve prx ztt, 14b-sml would have not contributed to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot- a manufacturing records evaluation (mre) was not performed as no lot number was retrieved for this device due to the condition (scratches) of the returned device.Corrected: h3.
|
| |
|