|
Correction: h3.Yes.H6.Investigation findings: 3251.Investigation conclusions: 22.Device evaluation: visual inspection confirms that the arm component has fractured.The distal fractured piece was not returned for investigation.The arm of the inserter is a subcomponent which features an ipsilateral prong that attaches to the interbody spacer for insertion.This type of damage occurs when force is applied during impaction/insertion of the interbody spacer causing the arm to detach from the shaft.Review of the device history records showed no manufacturing or processing related irregularities that would cause or contribute to this failure mode.Labeling review: warnings/cautions/precautions: potential risks identified with the use of these fusion devices, which may require additional surgery, include device component failure, loss of fixation, pseudoarthrosis (i.E.Non-union), fracture of the vertebra, neurological injury, and/or vascular or visceral injury.Possible adverse effects: initial or delayed loosening, bending, dislocation, and/or breakage of device components.
|