• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH EXPEDIUM SPINE SYSTEM SINGLE INNER SET SCREW 5.5; FIXATION, SPINAL INTERVERTEBRAL BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH EXPEDIUM SPINE SYSTEM SINGLE INNER SET SCREW 5.5; FIXATION, SPINAL INTERVERTEBRAL BODY Back to Search Results
Catalog Number 179702000
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported that the patient underwent a tirf for an iliac.During surgery, the si set screw was stripped, preventing it from turning and preventing the application of torque during final tightening.The sai screw f8.0 x 80 mm had stripped threads on the screw head, causing the set screw to be stripped.The problem was resolved by swapping the screw and the set screw.The surgery was completed successfully within 30 minutes of delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM SPINE SYSTEM SINGLE INNER SET SCREW 5.5
Common Device Name
FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key23075057
Report Number1526439-2025-01194
Device Sequence Number20030257
Product Code NKB
UDI-Device Identifier10705034064307
UDI-Public(01)10705034064307
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200245
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number179702000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/01/2025
Supplement Date Received by Manufacturer10/04/2025
Initial Report FDA Received Date09/17/2025
Supplement Report FDA Received Date10/06/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-