• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: 3M COMPANY 3M¿ RANGER¿ BLOOD/FLUID WARMING HIGH FLOW SET; WARMER, BLOOD, NONELECTROMAGNETIC RADIATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

3M COMPANY 3M¿ RANGER¿ BLOOD/FLUID WARMING HIGH FLOW SET; WARMER, BLOOD, NONELECTROMAGNETIC RADIATION Back to Search Results
Model Number 24355
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time, the product has not yet been returned for this complaint.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs (device history review) for the libre sensor and sensor kit was reviewed and the dhrs showed the libre sensor and sensor kit met specifications prior to release to distribution.If the product is returned, a physical investigation will be performed and a follow-up report will be submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
A user facility reported three 3m¿ ranger¿ blood/fluid warming high flow sets leaked while using a rapid infuser.There was a code transfusion in the intensive care unit and two different sets of tubing separated at the spike and tube connection, resulting in two staff members getting blood on their face.This also occurred during a code transfusion in the emergency department earlier in the week, resulting in exposure to a staff member.No injuries or medical interventions were indicated due to the alleged malfunctions.
 
Additional Manufacturer Narrative
Unused sets from the same lot were received for analysis.The reported issue was not confirmed, all samples passed specifications.The root cause of the alleged malfunction could not be determined.Possible causes include pulling on tubing instead of holding on to the spike, excessive twisting of spike, excessive force on tubing when removing spike from infusion bag, or over pressurization of set above 300 mmhg.A review of the batch record showed this lot met all specification for product release and no deviations were found that could have caused or contributed to the reported malfunction.Solventum will continue to monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3M¿ RANGER¿ BLOOD/FLUID WARMING HIGH FLOW SET
Common Device Name
WARMER, BLOOD, NONELECTROMAGNETIC RADIATION
Manufacturer (Section D)
3M COMPANY
2510 conway ave
st. paul MN 55144
Manufacturer (Section G)
NPA DE MEXICO S. DE R.L. DE C.V.
sor juana ines de la cruz #201
50 interior 5, cd. industrial
tijuana,baja california
Manufacturer Contact
yvonne mcmann
6203 farinon drive
san antonio, TX 78249
2103743188
MDR Report Key23075240
Report Number2110898-2025-00060
Device Sequence Number15891758
Product Code BSB
UDI-Device Identifier10707387792685
UDI-Public0110707387792685112502251728022510HX9911
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K973741
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24355
Device Catalogue Number24355
Device Lot NumberHX9911
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2025
Supplement Date Received by Manufacturer04/15/2026
Initial Report FDA Received Date09/17/2025
Supplement Report FDA Received Date04/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/25/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-