ETHICON INC. SURGICEL POWDER USA - ABSORBABLE HEMOSTA; AGENT, ABSORBABLE HEMOSTATIC, NON-COLLAGEN BASED
|
Back to Search Results |
|
| Catalog Number |
3013SP |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Wound Dehiscence (1154); Seroma (2069); Hypervolemia (2664); Insufficient Information (4580)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # = >(b)(4).D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H6.Component code: g07002 - device not returned to date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.1.Please provide the following patient demographic information, if available: age, gender, weight, bmi at the time of index procedure? 2.What was the date of index surgical procedure? 3.What were the diagnosis and indication for the index surgical procedure? 4.Please provide any relevant patient history: patient pre-existing medical conditions (i.E.Allergies, history of reactions), all concomitant medications, past medical history, any treatment required for events, dose, frequency, and therapy dates.5.Was there any intraoperative concurrent use of other products? 6.What is the lot number? 7.What was the intended use of the surgicel? was it used to address active bleeding or used prophylactically? 8.Where was the surgicel used (on what tissue)? 9.How much surgicel was used during the procedure? 10.Was the surgicel product left in place? was the excess irrigated and removed? 11.What were current symptoms following the index surgical procedure? what was the onset date? 12.Were cultures performed? if yes, results? 13.Has any surgical or medical intervention been performed? 14.What is physician¿s opinion as to the cause of this event? 15.Was there an alleged deficiency of the surgicel that contributed to the patient¿s post-operative seroma and dehiscence? 16.What is the patient¿s current status? 17.What is the users experience w/ surgicel powder and other hemostatic agents? this report is being submitted pursuant to the provisions of 21 cfr part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a pacemaker surgery procedure on an unknown date and absorbable hemostat was used.Post operatively the patient experienced seroma and dehiscence.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) additional information: a 2.Patient age at the time of event, a 2.Age unit, a 3a.Sex, h 6.Health effect - clinical code, h 6.Health effect - impact code.Additional information was requested, and the following was obtained: 1.Please provide the following patient demographic information, if available: age, gender, weight, bmi at the time of index procedure? male 52 2.What was the date of index surgical procedure? around (b)(6) 2025 for bring back procedure, original surgery a few days before.3.What were the diagnosis and indication for the index surgical procedure? ppm 4.Please provide any relevant patient history: patient pre-existing medical conditions (i.E.Allergies, history of reactions), all concomitant medications, past medical history, any treatment required for events, dose, frequency, and therapy dates.N/a 5.Was there any intraoperative concurrent use of other products? no 6.What is the lot number? not available 7.What was the intended use of the surgicel? was it used to address active bleeding or used prophylactically? active bleeding and prophylactically 8.Where was the surgicel used (on what tissue)? ep lab pacemaker pocket 9.How much surgicel was used during the procedure? entire 3g 10.Was the surgicel product left in place? was the excess irrigated and removed? yes, not irrigated.11.What were current symptoms following the index surgical procedure? what was the onset date? pressure and fluid buildup in wound.12.Were cultures performed? if yes, results? yes, negative 13.Has any surgical or medical intervention been performed? had to reopen, flush out wound and reclose.14.What is physician¿s opinion as to the cause of this event? doesnt know, blamed powder and stratafix.15.Was there an alleged deficiency of the surgicel that contributed to the patient¿s post-operative seroma and dehiscence? no 16.What is the patient¿s current status? - okay 17.What is the users experience w/ surgicel powder and other hemostatic agents? - used arista didn¿t irrigate and now needs to understand irrigation when using powder.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.1.Please provide the following patient demographic information, if available: age, gender, weight, bmi of each patient at the time of index procedure? kindly specify patient 1 (b)(4), patient 2 (b)(4) and patient 3 (b)(4).2.Please clarify if each procedure occurred on (b)(6) 2025.If not, please clarify the date of each patient¿s respective procedure; patient 1 (b)(4), patient 2 (b)(4) and patient 3 (b)(4).3.What is the product code of the stratafix suture used in each procedure (patient 1 (b)(4), patient 2 (b)(4) and patient 3 (b)(4)? 4.What was the initial approach for the index surgical procedure? (open, laparoscopic or other)? 5.On what tissue was the suture used? 6.What was the tissue condition (normal, thin, calcified, fragile, diseased)? 7.When beginning the running suture, did the surgeon take the first pass above or adjacent to the incision in a direction away from the incision? 8.After taking the first pass away from the incision, did the surgeon then take a pass of tissue perpendicular to the initial pass? if so, how many perpendicular passes? 9.Did the surgeon then proceed with wound closure in the opposite direction of the first pass? 10.Was the fixation tab seated against the tissue at the initiation of suture use during the index procedure? 11.Were two reverse stitches performed across the incision prior to closure? 12.Did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? 13.Did the stratafix suture break? if so, where was the break noted (termination, middle, end)? 14.Can you describe the appearance of the stratafix suture during second procedure, if applicable? 15.Can the surgeon please provide more detail on how they believe the surgicel powder and stratafix contributed to the event? 16.Were any pre-op cleansing procedures changed recently? if yes, please describe this report is being submitted pursuant to the provisions of 21 cfr part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) additional information was requested, and the following was obtained: 1.Please provide the following patient demographic information, if available: age, gender, weight, bmi of each patient at the time of index procedure? kindly specify patient 1 (b)(4), patient 2 (b)(4) and patient 3 (b)(4).2 male, last patient unknown.2.Please clarify if each procedure occurred on (b)(6) 2025.If not, please clarify the date of each patient¿s respective procedure; patient 1 (b)(4), patient 2 (b)(4) and patient 3 (b)(4) - surgeon didn't know, that week he said.3.What is the product code of the stratafix suture used in each procedure (patient 1 (b)(4), patient 2 (b)(4) and patient 3 (b)(4)? 2-0 spiral, 3-0 & 4-0 monocryl.4.What was the initial approach for the index surgical procedure? (open, laparoscopic or other)? open.5.What tissue was the suture used? subcu and skin.6.What was the tissue condition (normal, thin, calcified, fragile, diseased)? thin - ep lab 7.Was the fixation tab seated against the tissue at the initiation of suture use during the index procedure? not used, spiral only.8.Were two reverse stitches performed across the incision prior to closure? one stitch 9.Did the stratafix suture break? if so, where was the break noted (termination, middle, end)? middle.10.Can you describe the appearance of the stratafix suture during second procedure, if applicable? stratafix was intact, overall think it was due to surgicel not being irrigated and pressure building.11.Can the surgeon please provide more detail on how they believe the surgicel powder and stratafix contributed to the event? surgeon didnt irrigate surgicel and no active oozing, used prophylacticly - didn't listen to education or inservicing.12.Were any pre-op cleansing procedures changed recently? if yes, please describe ¿ no.The following information was requested, but unavailable: 1.When beginning the running suture, did the surgeon take the first pass above or adjacent to the incision in a direction away from the incision? unknown 2.After taking the first pass away from the incision, did the surgeon then take a pass of tissue perpendicular to the initial pass? if so, how many perpendicular passes? unknown 3.Did the surgeon then proceed with wound closure in the opposite direction of the first pass? unknown.4.Did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? this report is being submitted pursuant to the provisions of 21 cfr part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|