AIZU OLYMPUS CO., LTD. EVIS EXERA II COLONOVIDEOSCOPE Back to Search Results
Model Number CF-H180AI
Medical Device Problem Code Device Reprocessing Problem (1091)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause of the foreign material found within the air/water tube could not be further identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the colonovideoscope exhibited foreign matter in the air/water tube.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II COLONOVIDEOSCOPE
Common Device Name
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key23075768
Report Number9610595-2025-21234
Device Sequence Number4785560
Product Code FDF
UDI-Device Identifier04953170339585
UDI-Public04953170339585
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K100584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCF-H180AI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/27/2025
Initial Report FDA Received Date09/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1