|
An angiodynamics territory manager reported an issue with a 15 cm ire single activation electrode.During the 3rd ablation of a desmoid treatment, one probe started arcing, past the skin, while still inserted in the patient.The previous 2 ablations done with this device were successful with no issues.This only occurred while that probe was in active use, and was quickly shut down after just a few pulses.The physician then removed that probe and completed the ablation with the unaffected probes.The patient did not experience any adverse effects, harm, or require medical intervention because of this incident.
|
|
Correction: d2a the customer's reported complaint of probe sparked was able to be confirmed during evaluation of the returned probe; insulation was damaged.No manufacturing non-conformance was observed, which supports the event description of probe was utilized for two (2) ablation cycles prior to the sparking issue.Engineering evaluation: confirmed reported failure.Connected the device to a hypot tester and the probe failed every time trying new positions.There is damage to the outer insulation close to the tip that is allowing the device to arc.There was some kind of damage caused to the insulation.It is not possible to determine when the damage was caused.After assembly all devices are hypot tested.This defect would be caught during that testing.This device likely was damaged from the end to 3rd ablation of the desmoid treatment.The device would have arced during the first two ablations.Potential root cause for the electrical arcs (spark) observed is use of clamps on the probe needle.Dfu for the nk probes states the following: "do not attach anything to the device unless it is supplied by angiodynamics and indicated for use with this device.Attachments may damage the insulation and contribute to patient injury." a review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the directions for use (dfu) that is supplied with the nanoknife probe contains the following statements: warnings: do not use a device with damaged insulation.Do not attach anything to the device unless it is supplied by angiodynamics and indicated for use with this device.Attachments may damage the insulation and contribute to patient injury.Avoid having the nanoknife single electrode probes touch when delivering pulses.Ensure that both nanoknife single electrode probe electrodes are fully embedded within the targeted tissue prior to initiating pulse delivery.Do not attach anything to the device unless it is supplied by angiodynamics and indicated for use with this device.Attachments may damage the insulation and contribute to patient injury.Placing a single electrode probe without continual image guidance may increase the risk of unintended mechanical perforation, damage to critical anatomical structures, and/or unintended tissue destruction.Imaging equipment with high-definition capabilities is recommended for procedures where a target area involves or is adjacent to critical structures.Having inadequate imaging equipment may result in unintended mechanical perforation, damage to critical anatomical structures, and/or hemorrhage.Potential adverse effects adverse effects that may be associated with the use of the nanoknife system include, but are not limited to: arrhythmia, atrial fibrillation or flutter, tachycardia,fistula formation, damage to critical anatomical structure,(nerve, vessel, duct) hemorrhage, muscle contraction.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|