MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3507500BC
Medical Device Problem Codes Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 08/21/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: no information regarding explantation or an expected explantation date has been received.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a 55-year-old female patient underwent a primary breast augmentation with two 500cc mentor memorygel breast implants and experienced bilateral capsular contracture (baker grade 3) post-operatively, which was diagnosed with a physical exam.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.This report is for the right side device.
 
Additional Manufacturer Narrative
On december 28, 2025, the evaluation for the device was completed.Device evaluation summary: visual analysis of the returned device revealed that the breast implant had a tear on the anterior view, measuring approximately 2.1 cm.In addition, an area of shell abrasion was noted on the edges of the rupture.These findings are consistent with normal wear of the device.Shell abrasion suggests in-vivo folding or creasing of the device, which may be caused by continuous and sustained stresses to the device, such as an excessively small breast pocket or the folding or wrinkling of the shell within the breast pocket.In some cases, breast implants may also experience normal wear over time.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet as part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On october 27, 2025, mentor received additional information reporting that the patient underwent device explantation on (b)(6) 2025.It was also reported that the patient experienced a rupture on the right side.Code a0412 has been added in section h6-medical device problem code to capture this additional information.On october 14, 2025, mentor received additional information regarding the patient's experience.It was reported that the patient also experienced breast pain, inflammation, and mammary ptosis.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On november 27, 2025, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: capsular contracture and rupture.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key23076675
Report Number1645337-2025-10321
Device Sequence Number13278044
Product Code FTR
UDI-Device Identifier00081317000150
UDI-Public(01)00081317000150(17)121108(10)5785391
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/08/2012
Device Catalogue Number3507500BC
Device Lot Number5785391
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/25/2025
Supplement Date Received by Manufacturer10/27/2025
11/27/2025
12/28/2025
Initial Report FDA Received Date09/17/2025
Supplement Report FDA Received Date11/17/2025
12/15/2025
12/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age55 YR
Patient SexFemale
-
-