BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. INSYTE AUTOGUARD; PERIPHERAL IV CATHETERS Back to Search Results
Catalog Number 38183414
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that device did not activate safety feature when performing the puncture on the patient, the safety device did not activate, making it impossible to ensure the puncture, resulting in loss of venous puncture.
 
Additional Manufacturer Narrative
E1.All demographic information on the regulatory document states "confidential".H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event or Problem Description
No additional information.
 
Additional Manufacturer Narrative
A device history record review was completed for provided material number 38183414 and lot number 5091414.The review did not reveal any detected abnormalities during the production process that could have contributed to the reported defect and all quality tests were found to be within specification.As neither physical samples nor picture samples were available for return, a thorough sample analysis could not be completed.Based on the investigation results, an exact cause could not be determined for the reported event.However, the probable root cause could be related to spring formation during manufacturing.The operation and maintenance corrections performed various adjustments and alignments that may have been related to this defect.In addition, the reported issue of needle not activated mostly likely resulted from cylinder damage, possibly caused by pressure variation or a failure in the sensor reading during manufacturing.At this time, further action has not been determined necessary.Our quality team will continue to closely monitor the manufacturing process for signs of this potential defect and any emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSYTE AUTOGUARD
Common Device Name
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
BR 
Manufacturer (Section G)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
BR  
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key23077459
Report Number9610048-2025-00123
Device Sequence Number9812152
Product Code FOZ
UDI-Device Identifier00382903818341
UDI-Public(01)00382903818341
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number38183414
Device Lot Number5091414
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2025
Supplement Date Received by Manufacturer10/06/2025
Initial Report FDA Received Date09/17/2025
Supplement Report FDA Received Date10/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-