MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507320MC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
07/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.However, a photo of the suspect medical device was provided.An evaluation using the image was completed.Upon visual evaluation of the image provided in the complaint, the implant was observed ruptured.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation revision surgery with implantation of a 320cc mentor memorygel breast implant prosthesis.Post-operatively, the patient was diagnosed with a ruptured left implant using ultrasound imaging.As a result, the patient underwent left-sided exchange with a mentor gel implant on (b)(6) 2025.
|
| |
|
Additional Manufacturer Narrative
|
|
On december 03, 2025, the mentor analysis lab received the suspect medical device for evaluation.Paperwork received with the device provided a patient identifier.Patient identifier: (b)(6).On december 30, 2025, mentor completed an evaluation on the returned device.Device evaluation: mentor conducted visual inspection and leak testing of the device.During visual evaluation no apparent damage or visual anomalies were observed on the returned device.Leak testing was performed, according to mentor procedure, and it identified a tear on the posterior view, measuring approximately 0.9 cm.In addition, an area of shell abrasion was noted on the edges of the rupture.These findings are consistent with normal wear of the device.Shell abrasion suggests in-vivo folding or creasing of the device, which may be caused by continuous and sustained stresses to the device, such as an excessively small breast pocket or the folding or wrinkling of the shell within the breast pocket.In some cases, breast implants may also experience normal wear over time.Manufacturer¿s reference number: (b)(4).
|
| |
|