| Model Number |
VTICM5 12.6 (-12.50/1.50) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/06/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient lens rotation.The lens was exchanged with a same length but spherical lens.The problem was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Manufacturer narrative: rotation, secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: lens was returned in a micro-centrifuge vial, dry, with residue/debris on the lens.Visual inspection found no visible damage to the lens and fiber on the lens.Dimensional inspection found the lens within specifications.Claim#: (b)(4).
|
| |
|