The reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault, significant reduction of iridocorneal angles, significant shallowing of anterior chamber, blurred vision and glares/haloes.Due to excessive vault, significant reduction of corneal angles, significant shallowing of anterior chamber, blurred vision and glare/haloes, the lens was exchanged for a shorter length lens.The problem was resolved.Cause of the event was reported as patient related factor.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|