| Catalog Number |
8065000334 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Improper Flow or Infusion (2954)
|
| Health Effect - Clinical Codes |
Eye Injury (1845); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported a leaky valved cannula and a lack of air entering the eye, which resulted in the patient's eye starting to collapse with choroidal during vitrectomy surgery.The procedure was completed after replacing the replace all the infusion tubing up to the cassette, and then back into the eye under air and air began to flow normally at that time.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information updated in b 5, b 2.H.2 h.6.And h.11.Correction h.6 supplemental medical device report (smdr1) is being filed to correct the h.6, codes a1405, e0819 and g04084 to a050401, e2403 and g04133.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
This report represents the leaky valve cannula (27g) associated with the record.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.6.And h.11.No technical inquiries were received from the customer.A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.An acceptance quality inspection is performed to ensure product meets release acceptance criteria.This inspection includes evaluation for damaged valves.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|