ALCON RESEARCH, LLC - HOUSTON UNITY TOTAL PLUS POSTERIOR PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065000334
Medical Device Problem Codes Fluid/Blood Leak (1250); Improper Flow or Infusion (2954)
Health Effect - Clinical Codes Eye Injury (1845); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported a leaky valved cannula and a lack of air entering the eye, which resulted in the patient's eye starting to collapse with choroidal during vitrectomy surgery.The procedure was completed after replacing the replace all the infusion tubing up to the cassette, and then back into the eye under air and air began to flow normally at that time.There was no patient impact.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Additional Manufacturer Narrative
Corrected information updated in b 5, b 2.H.2 h.6.And h.11.Correction h.6 supplemental medical device report (smdr1) is being filed to correct the h.6, codes a1405, e0819 and g04084 to a050401, e2403 and g04133.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
This report represents the leaky valve cannula (27g) associated with the record.
 
Additional Manufacturer Narrative
Additional information provided in h.6.And h.11.No technical inquiries were received from the customer.A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.An acceptance quality inspection is performed to ensure product meets release acceptance criteria.This inspection includes evaluation for damaged valves.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNITY TOTAL PLUS POSTERIOR PROCEDURE PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23082163
Report Number1644019-2025-03479
Device Sequence Number14693928
Product Code LRO
UDI-Device Identifier00380650003348
UDI-Public(01)00380650003348(17)270331(11)250422(10)17CP29
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065000334
Device Lot Number17CP29
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2025
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/20/2025
Supplement Date Received by Manufacturer09/29/2025
11/19/2025
Initial Report FDA Received Date09/18/2025
Supplement Report FDA Received Date10/24/2025
12/11/2025
Date Device Manufactured04/22/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNITY CATARACT PROCEDURE PACK.; UNITY TOTAL PLUS POSTERIOR PROCEDURE PACK.; UNITY TOTAL PLUS POSTERIOR PROCEDURE PACK.
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-