• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. AZURE XT SR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. AZURE XT SR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number W1SR01
Medical Device Problem Codes Failure to Interrogate (1332); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the patient that device stopped working.The implantable pulse generator (ipg) remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
2025-10-22 it was further reported that the ipg had a telemetry issue.
 
Additional Manufacturer Narrative
Correction:b5 and h2.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURE XT SR MRI SURESCAN
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key23082248
Report Number3004209178-2025-15811
Device Sequence Number11923489
Product Code NVZ
UDI-Device Identifier00643169634626
UDI-Public(01)00643169634626(17)200414
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P980035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/14/2020
Device Model NumberW1SR01
Device Catalogue NumberW1SR01
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2025
Supplement Date Received by Manufacturer10/22/2025
Initial Report FDA Received Date09/18/2025
Supplement Report FDA Received Date10/22/2025
Date Device Manufactured10/24/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-58 LEAD
Patient Age76 YR
Patient SexMale
-
-