• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO VILLALBA CAPSURE SENSE LEAD MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO VILLALBA CAPSURE SENSE LEAD MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 407452
Medical Device Problem Codes Under-Sensing (1661); Capturing Problem (2891)
Health Effect - Clinical Code Cardiac Arrest (1762)
Date of Event 11/27/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced cardiac arrest.It was reported that the right ventricular (rv) lead exhibited undersensing on ventricular tachycardia (vt) monitor episodes and an increase in thresholds.It was also reported that the right atrial (ra) lead exhibited high thresholds and far field r-wave (ffrw) oversensing on stored electrograms (egm) resulting in false classification of events as atrial tachycardia/atrial fibrillation (at/af).The patient was externally shocked and paced.The rv lead was reprogrammed and remains in use.The ra lead remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: w1dr01 ipg implanted on (b)(6) 2024.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE SENSE LEAD MRI SURESCAN
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key23082734
Report Number2649622-2025-25950
Device Sequence Number9812394
Product Code DTB
UDI-Device Identifier00643169590519
UDI-Public(01)00643169590519(17)181007
Combination Product (Y/N)Y
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P830061
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/07/2018
Device Model Number407452
Device Catalogue Number407452
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 08/29/2025
Initial Report FDA Received Date09/18/2025
Date Device Manufactured11/02/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
457445 LEAD
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age72 YR
Patient SexMale
Patient Weight69 KG
-
-