|
On 28th august 2025, it was reported by an arthrex's employee via an email that for 1 unit of ar-3638, knotless fibertak¿ with #2 suture (5-box), the implant holder tip broke during being deployed.Loose pieces were retrieved during the case.This was detected during a shoulder instability repair on (b)(6) 2025.The case was completed successfully by using another implant.
|
|
Additional information: g3, h3, h6.Complaint allegation is confirmed by the attached picture, which shows the tips of the device were broken off.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user-applied excessive mechanical forces, misaligned insertion; prying/leveraging the device during insertion.
|