• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO S. DE R.L. DE CV MELODY TRANSCATHETER PULMONARY VALVE; PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO S. DE R.L. DE CV MELODY TRANSCATHETER PULMONARY VALVE; PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELI Back to Search Results
Model Number PB1018
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Device Stenosis (4066); Intravalvular regurgitation (4068)
Health Effect - Clinical Codes Pulmonary Valve Stenosis (2024); Pulmonary Valve Insufficiency/ Regurgitation (4452)
Date of Event 08/20/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately eleven years and six months following the implant of this transcatheter pulmonary bioprosthetic valve, severe central pulmonary regurgitation and stenosis was observed.Subsequently, a second valve was implanted valve-in-valve.Per the physician, the valve was replaced due to normal somatic outgrowth and the valve experiencing normal valve deterioration.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MELODY TRANSCATHETER PULMONARY VALVE
Common Device Name
PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELI
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key23084985
Report Number9617601-2025-01496
Device Sequence Number10509282
Product Code NPV
UDI-Device Identifier00613994663962
UDI-Public(01)00613994663962(17)160114
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P140017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/14/2016
Device Model NumberPB1018
Device Catalogue NumberPB1018
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2025
Initial Report FDA Received Date09/18/2025
Date Device Manufactured01/29/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age18 YR
Patient SexMale
-
-