| Model Number |
SET-0021-1 |
| Medical Device Problem Codes |
Leak/Splash (1354); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The following has been reported: nurse reported: dead-ender on s port missing.Therefore, when the set was primed the medication spilled out of the s port onto the floor.A preliminary review identified the following issue: administration set leak (external part).An active infusion was not stopped.Reporting as a conservative measure.No adverse effects were reported.Additional information is needed to complete the investigation.
|
| |
|
Event or Problem Description
|
|
Findings: no sample or picture available for analysis.Customer complaint cannot be valid.Root cause: the secondary port is locked and not available for use in set-0021-25, the only line to be used is the primary.Based on all known information, there was no malfunction or mis-assembly of the set.The issue is attributed to customer training/product knowledge.Current process controls detection: 100% visual inspection.Code no.: set-0021-25.Batch no.: fa25e12237.Exception generated: the batch record fa25e12237 was reviewed.No exceptions were generated that could classify as a possible root cause of this defect.Quality test results: the finished good lot has passed all sampling acceptance criteria for all tests performed including in-process testing and product testing.
|
| |
|