• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FRESENIUS KABI USA LLC IVENIX, INC.; LVP PRIMARY ADMINISTRATION SET, SINGLE-INLET, LOW-SORBING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FRESENIUS KABI USA LLC IVENIX, INC.; LVP PRIMARY ADMINISTRATION SET, SINGLE-INLET, LOW-SORBING Back to Search Results
Model Number SET-0021-1
Medical Device Problem Codes Leak/Splash (1354); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
The following has been reported: nurse reported: dead-ender on s port missing.Therefore, when the set was primed the medication spilled out of the s port onto the floor.A preliminary review identified the following issue: administration set leak (external part).An active infusion was not stopped.Reporting as a conservative measure.No adverse effects were reported.Additional information is needed to complete the investigation.
 
Event or Problem Description
Findings: no sample or picture available for analysis.Customer complaint cannot be valid.Root cause: the secondary port is locked and not available for use in set-0021-25, the only line to be used is the primary.Based on all known information, there was no malfunction or mis-assembly of the set.The issue is attributed to customer training/product knowledge.Current process controls detection: 100% visual inspection.Code no.: set-0021-25.Batch no.: fa25e12237.Exception generated: the batch record fa25e12237 was reviewed.No exceptions were generated that could classify as a possible root cause of this defect.Quality test results: the finished good lot has passed all sampling acceptance criteria for all tests performed including in-process testing and product testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IVENIX, INC.
Common Device Name
LVP PRIMARY ADMINISTRATION SET, SINGLE-INLET, LOW-SORBING
Manufacturer (Section D)
FRESENIUS KABI USA LLC
50 high street
suite 50
north andover MA 01845
Manufacturer (Section G)
FRESENIUS KABI USA LLC
50 high street
suite 50
north andover MA 01845
Manufacturer Contact
rebecca mccandless
3 corporate drive
lake zurich, IL 60047
8475502913
MDR Report Key23087577
Report Number3014732157-2025-00991
Device Sequence Number10484562
Product Code FPA
UDI-Device Identifier00811505030108
UDI-Public00811505030108
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K183311
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberSET-0021-1
Device Catalogue NumberSET-0021-1
Device Lot NumberFA25E12237
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/18/2025
Supplement Report FDA Received Date11/03/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/12/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-