• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TRABECULAR METAL MODULAR SHELL POROUS WITH MULTI HOLES 50 MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TRABECULAR METAL MODULAR SHELL POROUS WITH MULTI HOLES 50 MM; PROSTHESIS, HIP Back to Search Results
Catalog Number 00620205020
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994)
Date of Event 03/31/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: medical product: liner standard 3.5 mm offset 36 mm i.D.For use with 50/52/54 mm o.D.Shells: catalog# 00630505036, lot# 66229500; bone screw self-tapping 4.5 mm dia.40 mm length: catalog# 00625004540, lot# j7329546; bone screw self-tapping 4.5 mm dia.35 mm length: catalog# 00625004535, lot# j7545291; bone screw self-tapping 4.5 mm dia.50 mm length: catalog# 00625004550, lot# 64157927, qty#2; 50mm x 10mm thickness acetabular augment: catalog# 00489405010, lot# 65378688.G2: foreign: germany.Visual examination of the provided pictures identified the explanted shell, liner, augment, screws, and head covered in bio-debris.The augment is confirmed to be fractured.The locking ring is sitting slightly outside of the groove.No further evaluation can be made.Device history record was reviewed and no discrepancies related to the reported event were found.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: fracture of the medial wall of the left acetabulum with mild medial implant migration and cup loosening as noted.Root cause was unable to be determined.Reported event was confirmed by review of provided radiographs and photos of the explanted devices.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that a patient underwent an initial total hip arthroplasty.Approximately ten (10) months post-implantation, the patient began to experience pain and the acetabular augment was found to be fractured.The patient underwent revision surgery to replace the affected component without reported complication.Due diligence is complete as multiple attempts have been made; all available information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRABECULAR METAL MODULAR SHELL POROUS WITH MULTI HOLES 50 MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23087983
Report Number0001822565-2025-03476
Device Sequence Number9796374
Product Code LPH
UDI-Device Identifier00889024118836
UDI-Public(01)00889024118836(17)300228(10)64553183
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K093561
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number00620205020
Device Lot Number64553183
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2025
Initial Report FDA Received Date09/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age70 YR
Patient SexFemale
Patient Weight97 KG
-
-