| Catalog Number |
CMC1422P |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3004788213-2025-00054.
|
| |
|
Event or Problem Description
|
|
It was reported that an roi-c anchoring plate was deformed and an roi-c cage fractured during surgery.A second plate and cage were used to complete the surgery and there was no patient impact; however, there was a 20 minute delay to the procedure.This is report two of two for this event.
|
| |
|
Event or Problem Description
|
|
It was reported that an roi-c anchoring plate was deformed and an roi-c cage fractured during surgery.A second plate and cage were used to complete the surgery and there was no patient impact; however, there was a 20 minute delay to the procedure.This is report two of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in d9 and h3.Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the cage is fractured.Root cause: this event could possibly be attributed to off-axis or excessive forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|