SYNTHES GMBH QUICK COUPLING FOR K-WIRES; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
511.791 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and the reported condition that the quick coupling device would not attach to the compact air drive device was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was determined that the device was frozen/will not move.It was further determined that the device failed pretest for check function of the adjusting sleeve and check the tool coupling.The assignable root cause was determined to be traced to the user, which is user error.
|
| |
|
Event or Problem Description
|
|
It was reported that the quick coupling device would not attach to the compact air drive device.During service and evaluation, it was determined that the device was frozen/will not run.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported; however, it was reported that the event occurred in august of 2025.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition that the quick coupling device would not attach to the compact air drive device was confirmed.An assessment was performed and it was determined that the device was frozen/will not move and the coupling could not engage the attachment.It was further determined that the device failed pretest for check function of the adjusting sleeve and check the tool coupling.The assignable root cause was determined to be traced to the user, which is user error.
|
| |
|