• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; FENESTRATED BIPOLAR FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; FENESTRATED BIPOLAR FORCEPS Back to Search Results
Model Number 420205-16
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of customer provided image was performed by an intuitive surgical inc., (isi) regulatory post market surveillance analyst.Following information was provided: the image was consistent with complaint of broken cable.The 8mm fenestrated bipolar forceps instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
 
Event or Problem Description
It was reported that after a da vinci-assisted surgical procedure, the 8mm fenestrated bipolar forceps instrument was observed to have a broken cable after it was pulled out.The procedure was completed with no reported injury.
 
Event or Problem Description
Refer to h11 for follow-up information.
 
Additional Manufacturer Narrative
The instrument involved in this complaint has been received and tested by failure analysis.The instrument was found to have a damaged grip cable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
FENESTRATED BIPOLAR FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
4085232100
MDR Report Key23092907
Report Number2955842-2025-38625
Device Sequence Number9809089
Product Code NAY
UDI-Device Identifier00886874111642
UDI-Public(01)00886874111642(10)K10241010
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 01/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420205-16
Device Catalogue Number420205
Device Lot NumberK10241010 470
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/31/2025
Supplement Date Received by Manufacturer12/17/2025
Initial Report FDA Received Date09/19/2025
Supplement Report FDA Received Date01/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES.
Patient SexUnknown
-
-