MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SYNCHRONY; Cochlear Implant Back to Search Results
Model Number Mi1200 SYNCHRONY
Medical Device Problem Codes Impedance Problem (2950); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The user reported persistent pain over an extended period, as well as intermittent disconnections between the implant and the external device.
 
Event or Problem Description
The user reported persistent pain over an extended period, as well as intermittent disconnections between the implant and the external device.The user has been reimplanted.
 
Additional Manufacturer Narrative
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.According to the information received, the recipient suffered from persistent pain.Pain is a known undesired side effects of otologic surgery.Further, in situ measurements whilst implanted showed one channel in status hi due to undetermined reasons.The recipient has been reimplanted and is doing well with the new device.This is a final report.Note: imdrf investigation findings code is c1902.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43()577885626
MDR Report Key23093667
Report Number9710014-2025-00869
Device Sequence Number13213726
Product Code MCM
UDI-Device Identifier09008737094003
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/28/2022
Device Model NumberMi1200 SYNCHRONY
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/02/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2025
Supplement Date Received by Manufacturer09/07/2025
Initial Report FDA Received Date09/19/2025
Supplement Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age20 YR
Patient SexFemale
-
-