| Model Number |
559445 |
| Medical Device Problem Code |
High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 5086mri52 lead; implant date: (b)(6) 2012.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable pulse generator (ipg) exhibited premature battery depletion due to high right atrial (ra) lead thresholds.The ipg was explanted and replaced.The ra lead was capped and replaced. no patient complications have been reported as a result of this event.
|
| |
|