U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX PRO TRIFOCAL IOL W/AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX PRO TRIFOCAL IOL W/AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number PXYAT0
Medical Device Problem Code Failure to Deliver (2338)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
A non-healthcare professional reported during intraocular lens (iol) implant procedure, the surgeon noticed that the lens would not advance.Additional information was requested.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information was provided in h.3., h.6., and h.11.The product was returned for analysis, and the reported complaint was observed.The device was returned in a blister tray inside the product carton.The lock-out assembly has been removed.Viscoelastic is observed in the device.The lens is present in the lens bay.The lever is moving freely, and the plunger is not advancing.The device is taken apart for further evaluation.The canister is fully punctured.Non-company lubricant is observed on the canister neck.No damage observed to o-rings.The device was reloaded and was reactivated with a new canister.The failure mode could not be replicated; the device activated with no issues.The root cause for the reported complaint is deemed to be manufacturing related (the faulty sub-assembly is supplied by company's vendor in bray).The product did not meet specifications.The plunger is not advancing when the lever is pressed.Based on the results from company product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX PRO TRIFOCAL IOL W/AUTONOME DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23094620
Report Number9612169-2025-01867
Device Sequence Number8766202
Product Code HQL
UDI-Device Identifier00380658493165
UDI-Public(01)00380658493165(17)300215(11)250217(10)25990821
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPXYAT0
Device Catalogue NumberPXYAT0.140
Device Lot Number25990821
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/26/2025
Supplement Date Received by Manufacturer09/30/2025
11/18/2025
Initial Report FDA Received Date09/19/2025
Supplement Report FDA Received Date10/23/2025
12/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/17/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VISCOAT OPHTHALMIC VISCOSURGICAL DEVICE
Patient SexUnknown
-
-