| Model Number |
Ø9X 30MM BC IF SCRW, VENTED |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.One unpackaged ar-4030c-09 biocomposite¿ interference screw was received for investigation.Visual inspection identified significant surface damage.The tip of the screw was fractured.Functional testing could not be performed due to the extent of the device's damage.The most likely cause for the reported failure includes: improper bone preparation, resulting in inadequate or uneven seating of the screw.Misaligned insertion, causing the screw to engage bone at an incorrect angle or trajectory.Prying or leveraging the device during insertion, with excessive mechanical stress, can lead to thread deformation or damage.
|
| |
|
Event or Problem Description
|
|
On 1st sep 2025, it was reported by an arthrex's employee via an email that for 1 unit of ar-4030c-09, fastthread¿ biocomposite¿ interference screw 9 x 30 mm, its anchor broke during insertion.This was detected during an anterior cruciate ligament reconstruction of the knee joint surgery on (b)(6) 2025.The procedure was completed successfully by using another new ar-4030c-09.There were no patient harm and no secondary procedure needed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|