SHIRAKAWA OLYMPUS CO., LTD. 12G-SDI CABLE 8.5M Back to Search Results
Model Number MAJ-2429
Medical Device Problem Codes Erratic or Intermittent Display (1182); No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the subject device had the monitor blip and the error to show color on monitor before use.The issue occurred during set up / inspection for use.There were no reports of patient harm.
 
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, h2, h3, h6, h11 the device was returned to olympus for inspection, and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: malfunction of the cable.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: malfunction of the cable.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information received from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12G-SDI CABLE 8.5M
Common Device Name
SDI CABLE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key23095095
Report Number3002808148-2025-15655
Device Sequence Number8766214
Product Code FET
UDI-Device Identifier04953170415630
UDI-Public04953170415630
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K222861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAJ-2429
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/20/2025
Supplement Date Received by Manufacturer11/05/2025
Initial Report FDA Received Date09/19/2025
Supplement Report FDA Received Date12/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OTV-S700, 7403403
Patient SexUnknown
-
-