MEDTRONIC PUERTO RICO OPERATIONS CO. PUMP MMT-1780KPK 670G PATHWAY BLACK MG; AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-1780KPK
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Select patient information cannot be provided due to regional privacy regulations.The p-cap does not lock properly into the reservoir compartment.The following were noted during visual inspection: minor scratched lcd window, scratched case, missing o-ring (reservoir tube), detached retainer, pillowing keypad overlay, keypad overlay texture damage, serial number label missing.Conclusion summary: unit was confirmed with cosmetic damage at retainer area medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported to medtronic minimed that the customer experienced cosmetic damage.The customer reported no adverse event.The event involved product(s) mmt-1780kpk.Troubleshooting was not performed.No harm requiring medical intervention was reported.The customer will discontinue the use of the pump and revert to a backup plan per healthcare professional instructions.Mmt-1780kpk was requested and customer response was the device returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP MMT-1780KPK 670G PATHWAY BLACK MG
Common Device Name
AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
lusine boyadzhyan
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
MDR Report Key23095113
Report Number2032227-2025-263772
Device Sequence Number9503872
Product Code OZP
UDI-Device Identifier00643169939219
UDI-Public(01)00643169939219
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
P150001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-1780KPK
Device Catalogue NumberMMT-1780KPK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2025
Initial Report FDA Received Date09/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/16/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0956-2020
Patient Sequence Number1
Patient Age49 YR
Patient SexUnknown
-
-