C.R. BARD INC. (COVINGTON) -1018233 BARD® URINARY DRAINAGE BAG; DRAIN BAG Back to Search Results
Catalog Number 153504
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tube of the urine collection drain bag was deformed upon opening, and it was refrained from using it.Per follow up information received via rcc on 08sep2025, stated there was a slight kink in the inlet tube of the urine collection bag.
 
Additional Manufacturer Narrative
The complete initial reporter's facility name is (b)(6) hospital.The reported event was unconfirmed because the reported failure could not be reproduced.The returned device was functionally tested, and the product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.The labeling and instructions for use were found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Therefore, no additional action is required at this time.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® URINARY DRAINAGE BAG
Common Device Name
DRAIN BAG
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key23095131
Report Number1018233-2025-08329
Device Sequence Number18447883
Product Code KNX
UDI-Device Identifier00801741029530
UDI-Public(01)00801741029530(11)250423(17)270628(10)MYJZ4734
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number153504
Device Lot NumberMYJZ4734
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2025
Initial Report FDA Received Date09/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-