• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 8300AB
Medical Device Problem Code Device Stenosis (4066)
Health Effect - Clinical Codes Dyspnea (1816); Fatigue (1849); Insufficient Information (4580)
Date of Event 08/27/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated: b4, b5, b7, g3, g6, h2, h6 (health effect - clinical code, type of investigation, investigation findings, investigation conclusions).Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including hyperlipidemia, coronary artery disease, diabetes mellitus.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.
 
Event or Problem Description
It was reported that a patient with a 25mm 8300ab aortic valve underwent a valve-in-valve procedure after an implant duration of 7 years, 4 months due to stenosis (mg 41mmhg).The patient presented with shortness of breath and fatigue.The tavr procedure was successfully performed with a 26mm 9755rsl valve.The patient was in recovery post procedure and discharged on pod #2.
 
Event or Problem Description
It was reported that a patient with a 25mm 8300ab intuity valve underwent a valve-in-valve procedure after an implant duration of 7 years, 4 months due to aortic stenosis.The tavr was performed with a 26mm 9755rsl transcatheter valve.The procedure was successful.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES, LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key23095522
Report Number2015691-2025-07712
Device Sequence Number21003944
Product Code LWR
UDI-Device Identifier00690103194500
UDI-Public(01)00690103194500(17)210718
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P150036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/18/2021
Device Model Number8300AB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2025
Supplement Date Received by Manufacturer10/02/2025
Initial Report FDA Received Date09/19/2025
Supplement Report FDA Received Date10/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/19/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Life Threatening;
Patient Age70 YR
Patient SexMale
-
-